Model 521 and 522 IEEE488 Communications Errors
Repeat Addressing
The Krohn-Hite (EDC) Models 521 and 522 sold after June 2005 take advantage of HS488. These units do not need to be readdressed for continuous reads or writes. You will need to readdress the 521 or 522 if you address another unit with a different address.
Units sold before June 2005 conform to IEEE-488.1 standards and require GPIB addressing before any GPIB activity. You should configure your GPIB driver to perform repeat addressing. Refer to your GPIB controller manual for information about reconfiguring your controller.
Most units sold after January 2000 can be updated to take advantage of the HS488 version. Contact your local distributor or the factory for more information.
Termination Method
The model 521/522 must receive an end of message terminator to respond to a command string. Valid terminators that the model 521/522 will recognize include <CR> <LF> and EOI.
Switch Maintenance for Models: 330, 1030, 4032, 4600, 4601, CR Series, E100RC, MV Series.
The decade and range switches of the Models referred to above are lubricated at the factory. It is recommended that these switches DO NOT receive any service during the first year of operation. If you feel there is a problem, consult the factory.
Note: Over-zealous, arbitrary, or unnecessary cleaning may damage the switch contacts, which should be maintained below 10 milliohms.
Cleaning: Do not arbitrarily clean the switches. In many instances, lubrication may be all that is required. Relubricate After Cleaning! Lubricating - DO NOT USE OIL! The use of a cleaner without lubricating could shorten the life of the switches to about two months.
Contamination From Silicone Based Lubricants: Silicone based cleaners/lubricants should not be used where electrical contacts and other electrical components are used. Silicone will react with plastics and atmospheric contamination to form insulating films and silicone carbide compounds. Over time, silicone will react with plastics, softening them, thereby breaking down the integrity of the component (switch, relay, etc.). Silicone-based lubricants may also have been used as a mold release agent or as a general cleaner for housings. Since silicone can migrate great distances, it is recommended to clean the silicone off the surfaces. DeoxIT and PreservIT will displace the silicone.
Do Not Use Pencil Erasers: Do not use an eraser - any type of eraser - to clean contacts. Erasers are highly abrasive and will remove the precious metal plating. Further, the glue in erasers leave behind a film that is extremely difficult to remove which can later cause intermittent operation. Rubbing an eraser back and forth across the contacts can be a potential static generator.
General Information & Precautions: When using any aerosol, the unit should be de-energized (turned off) and used in well ventilated areas.
Switch Contacts cleaner. DeoxIT is a fast-acting, deoxidizing solution that cleans, preserves, lubricates and improves conductivity on metal connectors and contacts. Switch Contacts lubrication. PreservIT seals, lubricates and preserves metal surfaces for protection from oxidation and contamination. For use on clean/new surfaces or those pre-cleaned with DeoxIT. PreservIT contains no cleaners or de-oxidizers. It provides long-lasting protection for newly manufactured components or those cleaned by ultrasonics, DeoxIT or other procedures. PreservIT is recommended as a final treatment for surfaces that are in constant motion and/or subject to higher degrees of atmospheric contamination.
Spray Products Directions - DeoxIT Cleaner: Ideal for general purpose applications. Contains petroleum naphtha as the carrier solvent. It is flammable but non-aggressive to almost all materials. Apply a short burst to metal surface and operate device to assist breaking up of oxide layers. For severely oxidized surfaces, wait a few minutes before operating. Then use lint-free applicators on accessible surfaces and wipe until surface appears clean. In inaccessible areas flush away oxides with DeoxIT D5. As a final step, spray a short burst of DeoxIT for protection. Includes precise, three-way adjustable valve for controlled dispensing with minimal waste and over-spray.
PreservIT P5 Lubricant: Ideal for general purpose applications. Contains petroleum naphtha as the carrier solvent. It is flammable but non-aggressive to almost all materials. After cleaning metal surface, apply a short burst of PreservIT. If accessible, wipe with a lint- free applicator. If discoloration appears, the metal surface is not clean - use DeoxIT first. As a final step, spray a short burst of PreservIT for protection. PreservIT P5 includes precise, three-way adjustable valve for controlled dispensing with minimal waste and over-spray.
"Rule of Thumb": If the measurement indicates more than 1mVp-p. of noise on any Krohn-Hite instrument, the operator should recheck his equipment and lash-up.
Because noise may appear in many forms, Krohn-Hite recommends the use of an oscilloscope to make the noise measurements. A high gain 50 µV/cm or better, differential pre-amp such as the 1A7A, or the 7A22 Textronix models, or equivalent, are well suited for this application. In an environment with excessive EMI levels, these tests should be performed in a screen room. This will prove the specs of the Krohn-Hite unit, and will, with a comparison test in the normal environment, permit calibration for radiated noise pickup on the test measurements. The noise test should not be made simultaneously with regulation and voltage accuracy test. The "pump back" currents from some measuring devices will seriously disturb noise measurements.
Differential inputs measurements are the most reliable. They will cancel out common mode, due to slight errors in lash-up.
The scope and the Krohn-Hite calibrator under test should be connected to adjacent power outlets on the same phase. A three wire ground is required. In the event the line does not have a ground, the scope and unit under test should have a separate, heavy wire chassis-to-chassis connection separate from the shield of the differential input leads. The lead used between the scope input and the source output should be a shield, twisted pair with the shield connected to the frame of the scope, and, to the ground lug adjacent to the output terminals of the Krohn-Hite source. Do not use the shield of the input cable as the chassis-to-chassis connection in place of line system ground. Use additional separate heavy wire.
If the Krohn-Hite instrument has remote sensing, be sure that the "output and "sense" terminals are bussed. Set the output on the instrument to zero on each voltage range. Observe that ripple and noise do not exceed specified values.
Reset Protocol and Initialization Problems with IEEE-488.1 and IEEE-488.2
From time-to-time some customers have had problems with the *RTS command (IEEE-488.2) when trying to initialize or reset a Krohn-Hite device in their systems, especially when the device is the first in the chain. The RESET protocol implemented in IEEE-488.2 initializes an entire system using 3 levels of reset. However, devices that only implement IEEE-488.1 reset commands, will experience problems using the *RTS command. Below are the three levels the *RTS command implements:
The first level of reset places the system bus into an idle condition. All devices are unaddressed and put into a serial poll idle state and the system controller becomes the controller-in-charge. This first level is referred to as bus initialization.
The second level of reset ensures that a device can be sent a PROGRAM MESSAGE.
The third level of reset initializes device specific functions.
All Krohn-Hite instruments conform to the IEEE-488.1 Standard, not IEEE-488.2 If a program is written using the IEEE-488.2 protocols for reset, the device may hang the system. To initialize or reset the Krohn-Hite instrument, send a DEVICE CLEAR or an IFC command separately, do not include the Krohn-Hite device when sending the *RTS command. The *RTS command, as part of its 3 levels of reset, sends a message exchange initialization or a universal reset which will most likely hang the system.
Please be aware of the following warning stated in the ANSI/IEEE Std 488.2-1987 standard.
"WARNING: Devices which do not comply with IEEE-488.2 may behave much differently from what is described in this protocol for message exchange and device initialization. A bus initialization, IFC, should behave identically on all IEEE-488.1 devices. The behavior of non-IEEE-488.2 devices should be investigated before doing message exchange initialization with them. A device initialization should NEVER be done to a device which has not implemented the *RST command."
If a high impedance or low current load is connected, short pin B to pin F and short pin C to pin E. If an appreciable current is desired, the remote sense capability should be used. The calibrator utilizes a 4 wire output, eliminating the effect of IR drops on the instrument's accuracy.
NOTE: One of the Configurations Shown below must be used.
Pins B, C, E and F must All Be Used
Figure 1

Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
Email: sales@krohn-hite.com
Coasin
Virrey Del Pino 4071
1430 Buenos Aires
Tel: 54-11-4552-3185
Fax: 54-11-4555-3321
E-Mail: coasin@coasin.satlink.net
Website: www.coasin.com.ar
Universal Electronik Import GmbH
A-1230 Wien, Anton-Freunschlag-Gasse 49
Tel: 43/1/54 51 588
Fax: 43/1/54 51 464
E-Mail: sales@uei-vienna.com
Website: www.uei.at
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
Ohmtronics B.V.
Luchthavenweg 81209
5657 EA Eindhoven
Tel: +31-40-2573148
Fax: +31-40-2573178
E-Mail: info@ohmtronics.nl
Web: www.ohmtronics.nl
Suplitec Ltda
Rua Sena Madureira 455
31340-000 Belo Horizonte – MG
PO Box 2103
Tel: 55-313-498-1177
Fax: 55-313-441-0841
Email: suportc@suplitec.com.br
Website: https://www.suplitec.com.br
TestForce Systems
E-Mail: sales@testforce.com
Website: www.testforce.com
Montreal
9450 Trans-Canada Hwy
St. Laurent QC H4S 1R7
Tel: (514) 856-0970
Fax: (514) 856-6983
Toronto
2-1795 Ironstone Manor
Pickering ON L1W 3W9
Tel: (905) 839-4290
Fax: (905) 839-0452
Ottawa
107-235 Menten Place,
Nepean ON K2H 9C1
Toll Free: (888) 880-6804 Ext. 7321
Tel: (613) 829-6859
Fax: (613) 829-5908
Calgary
111-7879 8th St. NE
Calgary AB T2E 8A2
Tel: (403) 247-3725
Fax: (403) 202-2016
Din Instrumentor Ltda
Avda, Holanda 2023
PO Box 335-22
Providencia, Santiago
Tel: 562-205-0100
Fax: 562-225-8139
E-Mail: instrum@dininstrumentos.cl
Website: www.dininstrumentos.cl
Campbell Scientific Measurement Technology (Beijing) Co., Ltd.
Room 1201, 12th Floor, Building 60,
No. 109 Jinghai Third Rd, Beijing Economic/Technological Develop Zone, Beijing, China
Main Phone: 8610.6561.0080
Main Fax: 8610.6561.0081
Email: info@campbellsci.com.cn
Website: www.krohn-hite.com.cn
Email: info@campbellsci.com.cn
Website: www.krohn-hite.com
Tectra d.o.o.
Licka 33
HR-10000 Zagreb
Tel: +385 1615 2606
Fax: +385 1615 2605
E-Mail: zagreb@tectra.hr
Website: www.tectra.hr
Contact: Ivo Novakovic
Tectra A.S.
Domkovska 2342/43
193 00 Praha 9
Tel: 42-02-8192-1650
Fax: 42-02-8192-1649
E-Mail: tectra@tectra.cz
Website: www.tectra.cz
Metric A/S
Roskildevej 8 – 10
DK-2620, Albertslund
Tel: 45-4371-6444
Fax: 45-4371-6433
E-Mail: metric@metric.dk
Website: www.metric.dk
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
Intotel Oy
PL 125, 02631 ESPOO (Kutojantie 4)
p. +358 9 521 300
f. +358 9 755 3110
E-Mail: intotel@intotel.fi
Website: www.intotel.fi
Made in Broke
1003 Rue Joseph
REMUET
Tel: 04-74-07-80-33
Fax: 04-74-07-80-37
contact@madeinbroke.com
Nucletron Technologies GmbH
Riesstraße 8
80992 Munich
Tel: 49-89-149002-19
Fax: 49-89-149002-11
E-Mail: info@nucletron.de
Website: www.nucletron.de
Prodigit Electronics Co., LTD.
Shen Zen Office
Tel: 86-755-26063392
Fax: 86-755-26063390
E-Mail: grance@prodigit.com.tw
Tectra Electronics Kft
Mikszath Kalman U. 105
1184 Budapest
Tel: 361-291-1643 or 361-291-2065
Fax: 361-291-1643
E-Mail: tectra@tectra.hu
Website: www.tectra.com
Electronic Enterprises (India) Pvt. Ltd.
OFFICES
MUMBAI
306, Nimesh Industrial Estate
Mulund (E), Mumbai – 400 081
Phone : (022) 2563 9904
Fax : (022) 25633960
Email : eemulund@eeipl.in
Hyderabad Branch Office
403, Taramandal, 5-9-13, Saifabad,
Hyderabad – 500 004
Phone : (040) 23243352,
23237214, 23237215
Fax : (040) 23240172
Email : eehyd@eeipl.in
New Delhi Branch Office
M-4/23, Model Town
IIIrd Main Road, New Delhi – 110 009
Phone : (011) 27124 915
Fax : (011) 2724 0436
Email : eedelhi@eeipl.in
Kota Branch Office
Karmbhoomi
129-A, Shakthi Nagar, Kota – 324 009
Phone : (0744) 2501 113
Fax : (0744) 2501 113
Email : eekota@eeipl.in
Kolkata Branch Office
76/B-9, Raj Mohan
Banerjee Road, Kolkata – 700 035
Phone: (033) 2578 7114
Fax: (033) 2577 0551
Email: eecal@eeipl.in
Bangalore Branch Office
No:12, 1st Floor, 3rd Main Road, Shivanagar, West of Chord Road
Bangalore 560 010
Phone : (080) 2338 0451
Fax : (080) 2338 9273
Email : eebng@eeipl.in
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
RDT Equipment and Systems LTD
22, Efal StreetPetach-Tikva 4951122
Tel.: +972-3-6450708
Fax: +972-3-6450723
Mobile: +972-50-6890021
Call center: +972-3-6450780
E-Mail: ravi@rdtest.co.il
Website: www.rdtest.co.il
Made in Broke
1003 Rue Joseph
REMUET
Tel: 04-74-07-80-33
Fax: 04-74-07-80-37
contact@madeinbroke.com
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
MNTECH1 Co. ,Ltd
Head Office
1202-105, 102 Singal-ro
Giheung-gu, Yongin-si
Gyeonggi-do, Korea 446-590
Tel : +82-31-891-0011
Fax : +82-31-891-0032
Mobile phone : +82-10-3550-8248
sunglee@mntech1.co.kr
O’Connors Pte Ltd
801, Lorong 7 Tao Payoh, WBL Building
O’Connor’s Annex Block
Singapore 319319
Tel: 65-6470-2000
Fax: 65-6470-3000
Email: enquiries@oconnors.wbl.co.com.sg
Website: www.oconnor.co
Supervy Sistemas, S.A. de C.V.
Circuito Bahamas 9-16
Col. Lomas Estrella-Iztapalapa
Apdo. Postal 150-076
Tel: 52-55-5608-5300
Fax: 52-55-5656-6156
E-Mail: supervy@prodigy.net.mx
Website: www.supervy.com.mx
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
Ohmtronics BV
Gelderdijk 32
5975 NT Sevenum
Tel: +31-40-2573148
Fax: +31-40-2573178
E-Mail: info@ohmtronics.nl
Web: www.ohmtronics.nl
Instrutek A/S
Skreppestad Naeringspark
N-3261 Larvik
Tel: 47-33-16-5700
Fax: 47-33-16-5701
E-Mail: post@instrutek.no
Website: www.instrutek.com
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
AVESCO MARKETING CORPORATION
PO Box 3531, Manila, Philippines
810 Aurora Boulevard corner Yale Street,
Cubao, Quezon City, Philippines 1109
TEL#: (632) 912-8881
FAX#: (632) 912-2352 / 912-2911 / 912-2999
Email: import@e-avesco.com, avesco@email.com.ph
Unitronex Poland Sp z o.o.
Ul. Grochowska 341 suite 68
03-822 Warsaw
Poland
tel: +48 22 631-2643
email: info@unitronex.pl
Website: www.unitronex.pl
Unitronex Corporation (USA)
2155 Shelby Drive Suite C
Sedona, AZ 86336
Tel: 928-862-4037
E-Mail: info@unitronex.com
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
O’Connors Pte Ltd
801, Lorong 7 Tao Payoh, WBL Building
O’Connor’s Annex Block
Singapore 319319
Tel: 65-6470-2000
Fax: 65-6470-3000
Contact Email: enquiries@oconnors.wbl.co.com.sg
Website: www.oconnor.co
Tectra S.R.O.
Pod Amfiteatrom 7
PO Box 163
SK-93401 Levice
Tel: 421-36-6314581
Fax: 421-36-6312188
E-Mail: tectra@tectra.sk
Website: www.tectra.com
Tectra d.o.o.
Stegne 21
SI-1000 Ljubljana
Tel: +386 15114612
Fax: +386 15114613
E-Mail: zlatko.matic@tectra.si
Website: www.tectra.hr
Contact: Zlatko Matic
Protea Electronics (PTY) Ltd.
23 Galaxy Avenue
Linbro Park, Sandton 2065
Postal Address
Private Bag X19
Bramley 2018
South Africa
+27 (0)11 719-5700
Made in Broke
1003 Rue Joseph
REMUET
Tel: 04-74-07-80-33
Fax: 04-74-07-80-37
contact@madeinbroke.com
Ferner Elektronik AB
Fabriksvagen 2
746 34 Balsta
Tel: 46-08-760-8360
Fax: 46-08-760-8341
E-Mail: info@ferner.se
Website: www.ferner.se
Tectron AG
Rosswiesstrasse 29
CH-8608 Bubikon
Tel: 41-55-243-3800
Fax: 41-55-243-3838
E-Mail: tectron@tectron.ch
Website: www.tectron.ch
Campbell Scientific Hong Kong Ltd
Room 3602, Level 36 Tower 1, Enterprise Square Five, 38 Wang Chiu Road
Kowloon Bay, KL, Hong Kong
Main Phone: 8610.6561.0080
Tel: 886-2-29182620
Fax: 866-2-29129870
Email: info@campbellsci.com.cn
Website: www.krohn-hite.com.cn
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
Gama Ölçüm Teknolojileri Dış Ticaret Ltd.
Ankara Caddesi
Orta Mahallesi No 2A
Pendik İstanbul
Tel: 0212 – 320 48 22
Fax: 0212 – 320 48 22
GSM: 0542 – 616 25 29
EMail: info@gamateknik.com
Web: www.gamateknik.com
Krohn-Hite Corporation
15 Jonathan Drive, Unit 4
Brockton, MA 02301-5566
Tel: 508-580-1660
Fax: 508-583-8989
sales@krohn-hite.com
Caltest Instruments Ltd.
4 Riverside Business Centre
Walnut Tree Close
Guildford, GU1 4UG
Tel: 01483 302700
E-Mail: sales@caltest.co.uk
Website: www.caltest.co.uk
TECOTEC JSC
44/120 Truong Chinh
Hanoi, Vietnam
Tel: 84-4-576-3500
Fax: 84-4-576-3498
Email: www.tecotecjsc@hn.vnn.vn