10Hz to 10MHz
0.05º Accuracy
---------------------------------
Precision Phasemeter
Automatic Voltage Ranging
Automatic Ranging Display
Deviation Measurements
Related Models Model 6200B
- 0.5º Acc.; 1Hz to 1MHz; 0.1º Resolution Model 6600 - 0.5º Acc.; 10Hz to
10MHz; 0.1º Resolution
> Model 6610 - 0.05º Acc.; 10Hz to
10MHz; 0.01º Resolution Model 6620A
- 0.05º Acc.; GPIB; 10Hz to 10MHz; 0.01º Resolution
Features Accuracy:
Typically 0.02°
Resolution: 0.01°
Automatic Ranging Display
Automatic Voltage Ranging
Input Voltage Range: 10mV rms to 320V rms
Deviation Measurements
Accepts Any Waveform
Analog Output
Options Rack Mount Kit: Part no. RK-316 permits the installation of the
Model 6610 into a standard 19" rack spacing. 001: BCD Output. 002: 1Hz operation. 003: Rear panel BNC connectors for REFERENCE and SIGNAL inputs. Extended 1 Year Warranty: Part No. EW6610.
PERFORMANCE
The Model
6610 provides precision phase measurements with a typical accuracy of 0.02°
and 0.01° resolution over most of the frequency range. It will accept a wide
range of input signal levels from 10mV to 320V rms and input waveforms including
sine, triangle, square and pulses. A 5 digit, LED display provides continuous
direct readout of phase angles between 0.00° and 360.00° or ±180.00°. These
two ranges can be manually or automatically selected.
UNIQUE DESIGN
Two common factors will
affect the accuracy of a phasemeter are distortion and/or broadband noise
present on either or both the input signals. A unique circuit design used
in the 6610 significantly reduces any effects that may be caused by these
two factors.
Another problem
found in most phasemeters is the inability to respond to very small phase
angles, resulting in fluctuations or inconsistencies in the meter reading
(sometimes referred to as ambiguity). To overcome this problem, the 6610 uses
a design that eliminates any ambiguities that may occur at readings near 0°,
180° and 360°, and allows measurements as small as 0.01° to be made.
STANDARD FEATURES
The 6610 provides a RELATIVE
measurement mode which allows the monitoring of phase deviations without having
to make unwanted calculations. Also provided are an automatic selection of
the input voltage range, front panel indictors to indicate a too low/high
input voltage range, and an analog output for use with an external meter or
strip chart recorder.
APPLICATIONS
The 6610 is the ideal
instrument for precision phase angle measurements. Its superior performance
and affordable price, make it the best choice for a broad range of applications
which include monitoring servo control systems, low impedance measurements,
adjustment of crystal resonance, testing and adjustment of filter networks,
controlling laser trimming of resistor networks, measurements in ac power
systems, secondary standard and general testing and measurement.
SPECIFICATIONS
FREQUENCY RANGE:
10Hz to 10MHz (1Hz Optional)
PHASE MODE:
Absolute or Relative.
ACCURACY
Sinewave:
Input Volts (rms)
Frequency Range
Accuracy
10mV to 32mV
1Hz to 10kHz
0.2º
10mV to 100mV
10kHz to 50kHz
50kHz to 10MHz
0.35º
0.7º/100kHz
32mV to 100V
1Hz to 10Hz
10Hz to 10kHz
0.1º
0.05º
100mV to 100V
10kHz to 50kHz
50kHz to 100kHz
100kHz to 10MHz
0.05º
0.35º
0.35º/100kHz
100V to 320V
1Hz to 10Hz
10Hz to 5kHz
0.2º
0.1º
Squarewave: Double the sinewave specification.
Double
the sinewave specification.
INPUT Signal Amplitude:
Auto ranging from 0.01V rms to 320V rms. Waveforms: Sine, triangle, square and >50ns pulse. (The phasemeter
is triggered on the positive
going transition of the input waveform. A sinewave on the reference input
and a square wave on
the signal input is allowed). Impedance: 1 Megohm in parallel with a 50pF. Maximum DC Component: ±200 volts.
RESPONSE Time Constant:
>10Hz, less than 500ms; <10Hz, less than 5s. Settling Time: To within specified accuracy, within 1 to 8 seconds,
dependent on input amplitude
and frequency (>10Hz).
DRIFT Vs. Time (30 days
without CALIBRATE reset):
Sinewave:
±0.025° from 20Hz to 100kHz
±0.1° at 10Hz
±0.1° per 100kHz above 100kHz.
Squarewave:
±0.025° from 10Hz to 5kHz
±0.05° to 100kHz
±0.1° per 100kHz above 100kHz. Vs. Temperature (without CALIBRATE reset):
±0.01°/°C, 10Hz to 100kHz
±0.05°/°C to 1MHz.
±0.05°/°C per MHz above 1MHz.
GENERAL Analog Output:
(for use with an external meter or recorder) 0-3.6 or ±1.8 volts dc,
10mV dc/degree phase; Impedance 50 ohms. Display: 0.5", 7 segment, green LED. Display Ranges: Automatic, 0.00° to 360.00° or ±180.00°. Resolution: 0.01°. Repeatability: Better than 0.01°. Ambient Temperature Range: 0°C to 50°C. Front Panel Controls: POWER, METER RANGE Reference Waveform, Signal
Waveform,
REL PHASE, CALIBRATE, phase adjust (0°, 180° and 360°). Terminals (BNC):
Front Panel: REFERENCE and SIGNAL inputs.
Rear Panel: Analog Output, optional REFERENCE and SIGNAL inputs. Power Requirements: 90-132V or 198-264V, single phase, 50-400Hz,
40W. Dimensions and Weights: 3.5" (8.9cm) high, 16.5" (41.9cm) wide, 16"
(40.6cm) deep;
15 lbs (6.8kg) net, 18 lbs (8.1kg) shipping. Accessories: 3-terminal line cord; operating manual.
OPTIONS Rack Mount Kit:
Part no. RK-316 permits installation of the Model 6610 into a standard 19"
rack spacing. 001: BCD Output. 002: 1Hz operation. 003: Rear panel BNC connectors for REFERENCE and SIGNAL inputs. Extended 1 Year Warranty: Part No. EW6610.