Home  >  Specifications  >  Model 4601

Sign-Up to receive our e-mail newsletter.

Keyword Search


 


 

AC Voltage
------------------------------------
10mVac to 1000Vac / 2A Max.
------------------------------------
Standard/Calibrator
 


Features
Voltage Range:
10mVac to 1000Vac rms in 5 Ranges
Voltage Accuracy:
±0.05%

4-Internal Fixed Frequencies
External Operation 45Hz to 5kHz
All Solid State

Output Power:
up to 25VA or 2A (max.) Over-Load Protection

Very Low Output Impedance
Weight: 40 lbs (18.09kg)

Options
100 Ampere (AC) Calibrator, EDC Model 3210C
10 Ampere (AC) Calibrator, EDC Model 3200C
Phase Angle Control, EDC Model PAS-300.


DESCRIPTION
The Model 4601 AC Voltage Calibrator/Standard is designed for simplicity of operation and is intended as a “work horse” standard for laboratory, production line and field service environments.

There are 4-internal frequencies plus a setting to accept an externally supplied frequency. A set of lamps indicate the input levels of the external as well as the internal oscillators. In the event of an over-load or short-circuit condition, the instrument will trip and shut-down the output. A lamp will illuminate, indicating the “tripped” condition.

The controls for frequency, amplitude and range are well marked and easy to operate. The three external oscillator indicator lamps monitor the external oscillator’s input – when one is used.

APPLICATIONS
Uses for the Model 4601 as a calibrator include: AC Millivolt Meters, DVMs, Dynamometers and Accelerometers. It is also a reference for A/D Converters.

The Model 4601 can be used for the design and evaluation of amplifiers, converters, filters, rectifiers, regulators, demodulators, transformers, inductors, RMS converters, AC gain-control circuits and oscillators. It can also be used for testing and calibrating SCRs and SCR switching circuits. It is a power source for servo applications and as a reference/source to drive transconductance amplifiers.

SPECIFICATIONS

OUTPUT
Voltage Ranges (20% over-range):

Range

Settability

Resolution
(1ppm of
 range)

1000V

100V to 1211.110V

1mV

100V

10V to 121.1110V

100µV

10V

1V to 12.11110V

10µV

1V

100mV to 1.211110V

1µV

100mV

10mV to 121.1110mV

0.1µV

Voltage Accuracy (Limit-of-Error Method): All peripheral specifications, i.e., errors for line, load, temperature, drift and noise are listed for your information and are non-additive to the accuracy statement. The accuracies below are for all frequencies.

   1000V, 100V, 10V ranges ±(0.05% of setting + 0.005% of range)
   1V and 100mV ranges ±(0.05% of setting + 0.05% of range).

Voltage Stability (non-additive): 8 hrs, ±(0.0075% of setting+ 0.001% of range or 10µV;) 6 months, ±(0.015% of setting + 0.005% of range or 10µV).
Compliance Current (max, loading): All ranges except 100mV range, 25VA or 2 amperes maximum; 100mV range, minimal current (see Output Impedance).
Output Impedance: 1000V, 100V, 10V and 1V ranges, 1milliohm; 100mV range, constant 10 ohms.
Line Regulation (non-additive): 0.005%.
Load Regulation (non-additive): 0.0025%.
Overload Protection: Automatic trip, lighted front panel indicator, pushbutton reset.

FREQUENCY SPECIFICATIONS
Frequency Range: 45Hz to 5kHz†.
   Internal: 4 internal frequencies, factory installed 50Hz, 60Hz, 400Hz and 1kHz.
   External: This instrument accepts the frequencies within 45Hz to 5kHz† from an external oscillator.
   Input amplitudes between 3Vrms and 5Vrms which may be monitored by the “operate level” indicators.
Frequency Accuracy: ±0.01%.
Frequency Stability: ±0.001%/year.
Distortion: <0.5%. This is LIMIT of distortion throughout the entire band of the frequencies 45Hz to 5kHz at all voltage settings.
Amplitude Modulation: 0.02% rms max. at 59Hz to 61Hz.

GENERAL
Temperature: Operating limit, 0°C to 59°C, T.C. = 0.002%/°C; nominal operating, 20°C to 30°C.
Power Requirements: 333 watts, 105Vac to 125Vac, 60Hz or 220Vac to 240Vac, 50Hz.
Dimensions: 19" (48.26cm) wide, 7" (17.78cm) high, 13.75" (34.93cm) deep.
Weight: 48 lbs. (21.8kg) net; 53 lbs. (24.0kg) shipping.
Warranty: One year, includes accuracy and stability.

OPTIONS
100 Ampere (AC) Calibrator, EDC Model 3210C
10 Ampere (AC) Calibrator, EDC Model 3200C
Phase Angle Control, EDC Model PAS-300.

OPTIONS
Extended 1 Year Warranty: Part No. EW4601.

† The instrument is operational up to 5kHz. However, the amplitudes in the frequency ranges >1kHz to 5kHz
   are not certified. Certification at 5kHz is available at an additional charge.

Specifications subject to change without notice.

Copyright © February 23, 2006 Krohn-Hite Corporation. All rights reserved.


15 Jonathan Drive, Unit 4, Brockton, MA  02301-5566 U.S.A.
Tel: (508) 580-1660; Fax: (508) 583-8989; info@krohn-hite.com