GENERAL DESCRIPTION
The Model 4503 is a remotely controlled (GPIB) AC voltage reference system
for applications with
automatic test equipment. It is a highly versatile reference source, designed
to meet the needs of
computer system production line testing, ATE systems and standards laboratories.
The unit is also
calibrated with equipment traceable to the National Institute of Standards
& Technology. The 4503
is overload and short circuit protected and is fully operational in normal
environmental conditions.
SPECIFICATIONS
OUTPUT
Amplitude (True rms) Ranges and Resolution:
|
Ranges
(Plus 10%
Over-Range) |
Settability |
Resolution
(Discrete Steps) |
|
10mV |
1mV to 12.11110mV |
10nV |
|
100mV |
10mV to 121.1110mV |
100nV |
|
1V |
100mV to 1.211110V |
1µV |
|
10V |
1V to 12.11110V |
10µV |
|
100V |
10V to 121.11110V |
100µV |
Amplitude Accuracy:
|
f Band |
10mV, 100mV, 1V |
10V, 100V |
|
30Hz to 20kHz |
±(0.04% of setting
+ 0.004% of range + 25µV) |
±(0.04% of setting
+ 0.004% of range) |
|
20kHz to 40kHz |
±(0.1% of setting
+ 0.02% of range + 20µV) |
±(0.1% of setting
+ 0.02% of range) |
|
40kHz to 120kHz |
±(0.4% of setting
+ 0.06% of range + 35µV) |
±(0.20% of setting
+ 0.030% of range) |
Basis for Accuracy Statement: The statement
includes all peripheral, additive specifications
(errors for line and load changes, ambient temperature changes and stability are
included in the
accuracy statement and are designated “non-additive”). Thus, this instrument may
be used under
minimally controlled environment and unmonitored conditions.
NOTE: All specifications, including accuracy,
are valid and warranted for twelve months.
Harmonic Distortion and Noise (band-pass 10Hz
to 1MHz):
|
f Band
|
10mV to 10V |
100V |
|
30Hz < 50Hz |
0.5% of setting +
20µV |
0.6% of setting |
|
50Hz < 100Hz |
0.07% of setting
+ 20µV |
0.07% of setting |
|
100Hz < 20kHz |
0.06% of setting
+ 20µV |
0.06% of setting |
|
20kHz < 40kHz |
0.08% of setting
+ 20µV |
0.2% of setting |
|
40kHz < 60kHz |
0.2% of setting +
20µV |
0 0.4% of setting |
|
60kHz to 100kHz |
0.2% of setting +
20µV |
0.6% of setting |
Compliance Current (Loading): 10mV and 100mV
range, Zout = 10W (into loads >5kW);
1V, 10V and 100V range, 10mA.
Load Regulation (non-additive, no load to full load): 10mV and
100mV range, not applicable;
1V, 10V and 100V range, <10kHz 0.005%, >10kHz 0.01%.
Line Regulation (non-additive): ±0.001% of setting for 10% line
change.
Stability (non-additive): 24hrs., 0.005%; 1 month, 0.007%; 6 months,
0.01%.
Temperature Coefficient: Ambient, 0.005%/°C; operating limit, 0.007%/°C.
FREQUENCY SPECIFICATIONS
|
Range |
Settability |
Resolution (100ppm) |
|
100kHz |
10kHz to 111.10kHz |
10Hz |
|
10kHz |
1kHz to 11.110kHz |
1Hz |
|
1kHz |
100Hz to 1.1110kHz |
0.1Hz |
|
100Hz |
30Hz to 111.10Hz |
0.01Hz |
Frequency Accuracy (Crystal Controlled): 0.001%
of setting.
Frequency Stability: 24 hrs., 0.0005% of setting; 1 month, 0.0008%
of setting;
6 months, 0.001% of setting.
TTL Compatible (Squarewave) Output: This is an auxiliary output,
synchronous with the
sinewave output. Output will drive 10 TTL loads.
PROGRAMMING
Manual Control and Display:
All functions may be controlled manually (“local”). A single front
panel switch controls the “remote/local” mode without disconnecting the program
lines. The 4503
will continue to accept and store all up-dated program commands while set in the
“local” mode. The
displays indicate the status of the instrument in either the “local” or “remote”
modes depending on
the mode selected.
Remote Programming Subsets (IEEE-488.1, GPIB Interface): SH1, AH1,
T6, L4, SR1, RL0,
PP2, DC0, DT0 and E1.
Programming Format: Integrated microprocessor GPIB interface.
GENERAL SPECIFICATIONS
Power Requirements (selectable):
115Vac ±10%, 60Hz, 50 watts; 220Vac ±10%,
50Hz, 50 watts.
Temperature: Calibration, 23°C ±1°C; ambient, 20°C to 30°C; operating
limit, -10°C to 50°C;
storage, -40°C to 85°C.
Dimensions (rack and bench mountable): 19" (48.3cm) wide, 3.5"
(8.9cm) high,
18" (45.8cm) deep.
Weight: 20 lbs. (9.1kg) net; 28 lbs. (12.7kg) shipping.
Warranty: One year, parts and labor.
Calibration Cycle: 12 months.
ACCESSORIES (optional)
Hermaphroditic Connectors and Cable
Assembly: Part No. 3045-1, for GPIB interconnections.
Length, 1 meter or 3045-2 length 2 meters.
Extended 1 Year Warranty: Part
No. EW4503.
AUXILIARY INSTRUMENTS
Model PCS-2B, 10 Ampere
Current Standard: May be used in conjunction
with the Model 4503.
Specifications and prices are subject to change without
notice.
Copyright ©
February 23, 2006 Krohn-Hite Corporation.
All rights reserved.
15 Jonathan Drive, Unit 4, Brockton, MA 02301-5566 U.S.A.
Tel: (508) 580-1660; Fax: (508) 583-8989; info@krohn-hite.com