1000V DC
------------------------------------
±0.01µVdc to ±1100Vdc
------------------------------------
Standard/Calibrator



Features
Outputs: 0.01µV to 1100Vdc
Current: up to 50mA
Accuracy: 0.002%
Solid State Hybrid Design
4-Terminal Remote Sensing
Warm-Up in <30 Seconds - Traceable to N.I.S.T

Options
Extended 1 Year Warranty: Part No. EW1030.

Related Models
   Model 511 - ±10ppm 100nVdc to 10Vdc
 


GENERAL DESCRIPTION
The Model 1030 is a multi-range, high accuracy DC voltage source and calibrator with an absolute error of less than 10ppm.  It is laboratory calibrated against a primary measuring system that is traceable to the National Institute of Standards & Technology.  Controls are from the front panel switches and decade switches.  No trims of adjustment are required between calibration cycles.

APPLICATIONS
Applications for the Model 1030 are as a system reference for analog computers, digital converters, strain gauges, telemetry, data acquisition, testing or data handling using analog voltage inputs and as an ultra-stable power supply.

SPECIFICATIONS

OUTPUT

Range Full Scale Resolution
(0.1ppm)
Current
(with 4-terminal remote sensing
and current limit control)
Zo
1000V ±1111.1110Vdc 100µV 15mA 0.03 ohms
100V ±111.11110Vdc 10µV 50mA 0.03 ohms
10V ±11.111110Vdc 1µV 50mA 0.03 ohms
100mV ±111.1110mVdc 0.01µV into high impedance loads
3 ohms output impedance
3 ohms

Output Accuracy (2 Methods):
    Calibration Accuracy: ±(0.002% of setting + 0.0005% of range + 5µV); The calibration Accuracy is at
       standard reference conditions at time of calibration. Nominal calibration temperature of 23°C ±1°C at
       <70% relative humidity. Nominal line voltage 117Vac at constant external load.
    Limit-of-Error (worst case) Accuracy: ±(0.003% of setting+ 0.001% of range + 5µV); the Limit-of-Error
       accuracy includes peripheral, additive specifications (i.e., error for line and load changes, ambient
       operating temperature changes and stability); calibration cycle, one year.
Ripple & Noise, rms, 0.1Hz to 100kHz band-pass:

Range Ripple & Noise
1000V < 175µV
100V <100µV
10V < 75µV
100mV <10µV

Load Regulation (non-additive): ±(0.0005% of range).
Line Regulation (non-additive): ±0.0002% of range for a 10% line change.
Stability (non-additive): 8hrs., ±0.00075%; 1yr., ±0.003%.

GENERAL SPECIFICATIONS
Temperature:
Calibration, 23°C ±1°C; ambient, 20°C to 30°C; operating limit, –10°C to 50°C; storage,
–40°C to 85°C.
Temperature Coefficient: Ambient, ±0.0005%/°C; operating limit, ±0.001%/°C.
Common Mode Rejection: 120dB, dc to 400Hz.
Isolation: Either terminal may be floated 1000Vdc with respect to chassis ground; 1000M ohm, 20pF.
Power Requirements: 25 watts, 50Hz to 400Hz, 115Vac or 230Vac ±10%.
Dimensions: 5.22" (13.3cm) high, 17" (43.2cm) wide, 14.5" (36.8cm) deep.
Weight: 13 lbs. (6.0kg) net; 17 lbs. (7.7kg) shipping.
Circuit Condition Indicator: Front panel indicator for short-circuit, overload, over-voltage condition, low-line voltage and malfunction.
Protection: Short-circuit and overload protection; automatic recovery.