Home  >  Products  >  Source/Calibrators  >  Model 511

Sign-Up to receive our e-mail newsletter.

Keyword Search


 

 

±100nVdc to ±10Vdc
------------------------------------
±10ppm Absolute Accuracy
Max. Current: 100mA
------------------------------------
DC Voltage Source/Calibrator


Features

Absolute Accuracy: ±10ppm
Voltage Range:
±100nVdc to ±10Vdc
Resolution:
1ppm of Range
Wideband Noise:
<5µVrms

3 Voltage Ranges

Max. Current:
100mA
Impedance:
10µ ohm all ranges

Floating Output
Certified to N.I.S.T.

One Year Warranty

Options
Option 003: Provides both front and rear panel terminals for instrument outputs. Only one output set of terminals may be used at one time.
Option 004: Provide CE regulation safety terminals.
Rack Mount Kit: Part number RK-314 permits installation of the Model 511 into a 19” rack spacing.
Extended 1 Year Warranty: Part No. EX511

Related Models
> Model 511 - ±10ppm 100nVdc to 10Vdc
   Model 521/RA5 - ±20ppm 100nVdc to 1000Vdc

  
Model 522 - ±20ppm 100nVdc to 100Vdc
   Model 523 - ±4ppm 100nVdc to 100Vdc
   Model CR103 - 100nVdc to 10Vdc
 


DESCRIPTION

Continuing with Electronic Development’s (EDC) tradition, the NEW Krohn-Hite Model 511 Precision DC Source/Calibrator is a secondary standard that is highly stable and repeatable from ±100nVdc to ±10Vdc, providing N.I.S.T. traceable voltages for use in production, calibration labs, QA and QC departments, design labs, or any place where an accurate voltage is needed.

Using Krohn-Hite’s state-of-the-art reference, the 511 provides extremely accurate (±10ppm) and stable (±2ppm for 24 hours) voltages over 3 ranges, allowing the output to be set from ±100nVdc to ±11.11110Vdc with a resolution of 1ppm of range.

The Model 511 is an extremely quiet source with <5µV of noise measured from 0.1Hz to 100kHz. Maximum load current is 100mA on all ranges.

RANGES
The  511 also provides 3 voltage ranges, ±100mV, ±1.0V and ±10V allowing the user to set the output from ±100nVdc to ±11.11110Vdc.  Resolutions are 100nV, 1µV and 10µV respectively.

APPLICATIONS
The Model 511 can be used in many applications including the calibration of precision A/D converters, strip chart recorders and calibration of Kaye Instruments Data Logging Systems.

SPECIFICATIONS

OUTPUT
Specifications apply at 23ºC ±1ºC, <70% relative humidity.

Range

Full Scale

Resolution

Absolute Accuracy ±(ppm of setting + µV)

1 Year
100mVdc ±111.1110mVdc 100nVdc 10 + 2
1.0Vdc ±1.111110Vdc 1µVdc 10 + 6
10Vdc ±11.11110Vdc 10µVdc 10 + 42

 Above specifications include stability, line regulation and load regulation.

 

Noise and Ripple

Range 0.1Hz - 10Hz 10Hz - 100kHz Max. Load Zo
100mV 2µVp-p max. 5µVrms 100mA 10µ ohms
1V 2µVp-p max. 5µVrms
10V 2µVp-p max. 10µVrms

Above specifications include stability, line regulation and load regulation.

Temperature Coefficient, ±(ppm of setting + µV)/ºC:

0ºC to 10ºC

3 + 2

10ºC to 40ºC 2 + 1

Stability:
  
24hrs, ±(2ppm of setting + 1µV).
   1 year, ±(10ppm of setting + 2µV).
Settling Time:
100ms max., typically 60ms.
Load Regulation (all ranges):
±1µV for a change from no load to full load.
Line Regulation (all ranges):
±(1ppm + 1µV of range) for a ±10% change from normal line voltage.

GENERAL
Calibration Interval:
1 year for ±10ppm accuracy.
Warm-Up Time:
1 hour to rated accuracy.
Isolation:
Floating 500V from chassis.
Protection:
Will drive a short-circuit indefinitely.

Temperature:

Calibration

23ºC ±1ºC.

Operating

0ºC to 50ºC.
Accuracy is de-rated above 40ºC due to loss of oven control.

Storage

-30ºC to 70ºC.

Power Requirements: 105-125 or 210-250 volts ac, single phase, 50Hz-400Hz, 20 watts.
Dimensions:
3.5” (9cm) high, 14” (36cm) wide, 12.5” (32.1cm) deep.
Weight:
12 lbs (5.4kg) net; 14 lbs (6.3kg) shipping.
Controls:
Six decade rotary switches with indicator knobs from 0 to 10 and 10% over-range; 3 position range and polarity rotary switches; push-button power on, sense and ground switches.
Indicators:
LEDs indicate voltage range, power on and overload.
Terminals:
Output terminals are safety, low thermal, gold plated type, mounted on the front panel. Terminal sets are configured for remote sensing of the output as follows:

    High Output and High Sense
    Low Output and Low Sense
    Case Ground

Output terminals conform to European Standard Specifications.

OPTIONS
Option 003:
Provides both front and rear panel terminals for instrument outputs. Only one output set of terminals may be used at one time.
Option 004:
Provide CE regulation safety terminals.
Rack Mount Kit: Part number RK-314 permits installation of the Model 511 into a 19” rack spacing.
Extended 1 Year Warranty: Part No. EX511

OPTIONAL ACCESSORIES
CAB-005:
Cable, Two Conductor Shielded Balance Line
CAB-018:
Cable, Multi-stacking Double Banana plug
CAB-023:
Cable Set, Low Thermal EMF Retractable Banana
CAB-024:
Cable Set, Low Thermal EMF Spade Lug

Specifications are subject to change without notice.

Copyright © April 21, 2006 Krohn-Hite Corporation. All rights reserved.


15 Jonathan Drive, Unit 4, Brockton, MA  02301-5566 U.S.A.
Tel: (508) 580-1660; Fax: (508) 583-8989; info@krohn-hite.com